Can I test my electro surgery generator’s REM / ARM / CQM alarm function with the ESU-2050?
Can I test my electro surgery generator’s REM / ARM / CQM alarm function with the ESU-2050?
The ESU-2050, by design, does not have REM/ARM/CQM testing capability built in. Instead, BC Group offers the CQM-2000 module that has been specifically designed to test this vital patient safety feature in generators by all popular electro surgery generator manufacturers. The CQM-2000 has 0.5% tolerance resistors and goes from o to 210 ohms in 1-ohm steps to address the needs of specific generators from specific manufacturers.