What could cause an SA-2000 series ESA not to indicate an earth leakage current reading above 000?
Check the Fuse Verify Test Receptacle is On. (HOT Switch) Verify the UUT is turn On.
Check the Fuse Verify Test Receptacle is On. (HOT Switch) Verify the UUT is turn On.
Check the fuse. This is an indication of a Open/Blown 250mA/250V fuse, (P/N UF-0250-01) on the front panel.
Check the fuse. This is an indication of a blown 250mA/250V fuse, (P/N UF-0250-01) located on the front panel.
Check the fuse. If the fuse on the front of the Safety Analyzer is blown these symptoms can be present. Check for a blown or improper fuse before sending the equipment in for repair.