What could cause my SA-2000 Safety Analyzer resistance reads high compared to another Safety Analyzer? Check the chassis cable, it could be intermittent or defective. Check the receptacle, it could be an indication of worn receptacle....
Why does the Ground stay Open during the current leakage test? Check the Fuse, it is most likely Open/Blown Replacement Fuse for SA-2000 series: 250mA/250V fuse, (P/N UF-0250-01) located on the front panel.
How to prevent blowing fuses and relays in safety analyzers? This will hold true for any safety analyzer. Plug the UUT into the test receptacle. Select equipment current, earth leakage or enclosure leakage. Select forward...
A Bussman type FWH-1/2, 250V, 1/4×1-1/4″ BC Group P/N UF-0500-01.
A 250mA/250V, 5X20mm, Fast Acting fuse, (P/N UF-0250-01)